Home | » Products | » Bpap Gii |
Technical Specification :
Dimensions: | 170 × 196 × 118 mm | |
290 × 196 × 134 mm (with humidifier) | ||
Weight: | < 1.5 kg | |
< 2.5 kg (with humidifier) | ||
Product Use, Transport and Storage | ||
| Operation | Transport and Storage |
Temperature: | 5oC to 35oC (41oF to 95oF) | -25oC to 70oC (-13oF to 158oF) |
Humidity: | 15% to 93% Non-condensing | 15% to 93% Non-condensing |
Atmospheric Pressure: | 760 to 1060 hPa | 760 to 1060 hPa |
Mode of Operation | Continuous | |
SD Card | With a capacity = 4G, the SD card can record patient data and fault information. Furthermore, the language pack stored on the SD card enables you to change the language of the device. | |
AC Power Consumption | 100 - 240 V AC, 50/60 Hz, 2.0 A max | |
Type of Protection Against Electric Shock | Class II Equipment | |
Degree of Protection Against Electric Shock | Type BF Applied Part | |
Degree of Protection Against Ingress of Water | IP22 |
Mode Comparison :
Model | Product Contents | Work Mode | Maximum Work | ||
LCD Screen | Optional Accessory1 | Optional Accessory2 | |||
T-20S | 2.4 inches | Heated | SpO2 Kit | CPAP, S | 20 |
T-20A | 2.4 inches | CPAP, S, Auto | |||
T-20T | 2.4 inches | CPAP, S, T, S/T | |||
T-25S | 3.5 inches | CPAP, S | 25 | ||
T-25A | 3.5 inches | CPAP, S, Auto | |||
T-25T | 3.5 inches | CPAP, S, T, S/T | |||
T-30T | 3.5 inches | CPAP, S, T, S/T | 30 |
Pressure Range :
IPAP: | 4.0 ~ 20.0 cm H2O (only applies to T-20S, T-20A, T-20T); |
| 4.0 ~ 25.0 cm H2O (only applies to T-25S, T-25A, T-25T); |
| 4.0 ~ 30.0 cm H2O (only applies to T-30T) in 0.5 cm H2O increments. |
EPAP: | 4.0 ~ 20.0 cm H2O (only applies to T-20S, T-20A, T-20T); |
| 4.0 ~ 25.0 cm H2O (only applies to T-25S, T-25A, T-25T, T-30T); in 0.5 cm H2O increments. |
CPAP Mode | 4.0 ~ 20.0 cm H2O | |
| Under single fault conditions, < 30 cm H2O for CPAP mode, < 40 cm H2O for the rest modes. | |
Pressure Display Accuracy | + (0.8 hPa + 4%) | |
Static Pressure Stability | + 0.5 hPa | |
Ramp | The ramp time ranges from 0 to 60 minutes | |
Sound Pressure Level | <30 dB, when the device is working at the pressure of 10 hPa. | |
Sound Power Level | <38 dB, when the device is working at the pressure of 10 hPa. | |
SpO2 | Range: 0-100%The margin of error for SpO2 between 70% and 100% is +3%. No strict accuracy requirements for SpO2 below 70%. | |
Pulse Rate | Range: 40-240 bpm | Margin of Error: + 3% |
Wavelengths | Red: 663 nanometers | Infrared: 890 nanometers |
Maximal Optical Output Power | Less than 1.5 mw maximum average | |
Tube | Length: 6 ft. (1.83m) | |
Heater Settings | 1 to 5 (95 to 167oF/ 35 to 75oC) | |
Maximum Operating Pressure | 30 hPa | |
Pressure Drop with Humidifier | < 0.4 hPa at 60 LPM flow | |
Humidifier Performance | Humidity Output: No less than 10 mg H2O/L | Environmental Conditions: Maximum airflow, 35oC, 15% relative humidity |
Maximum Delivered Gas Temperature | < 43oC | |
The Form and the dimensions of the Patient Connection Port | The 22 mm conical air outlet complies with ISO 5356-1 |